RZBA_40_tmbRadiometer surface contamination of alpha-and beta-active substances RZBA-04-04M (the radiometer) is designed to measure contamination of surfaces (skin or clothing of staff) of beta-and alpha-active substances (in flux density of alpha and beta particles incident the input window of detection units) and signaling of excess (or exceeded) established thresholds.

Special placement of stationary detection units BDZB on the body of the radiometer allows for simultaneous measurement of 18 areas of surface contamination of the garment (or skin) of the human beta-active substances. Measurement of all the other sites provide remote detection unit BDZB (for measuring contamination by beta-active substances) and the remote detection unit BDZA (for measurement of alpha-active substances).

Radiometer RZBA-04-04M relates to measuring devices with a threshold device in which sound and light alarm is triggered when the preset threshold is exceeded the level of contamination by beta-and alpha-active substances.

Radiometer RZBA-04-04M produces a signal "CLEAN" and unlocks the device prohibition output, if any of the stationary channel was not exceeding the level of contamination by beta - the active substances over a given threshold.

Radiometer RZBA-04-04M produces a signal "DIRTY" in the event of excess pollution beta - active substances over a given threshold for at least one stationary channel.

Radiometer RZBA-04-04M has the ability to connect to a local area network (LAN) through an interface RS485 (or optional, RS232).

Detection units BDZB and BDZA provide flow transformation of beta-and alpha-particles, respectively, statistically distributed in a sequence of voltage pulses, whose frequency depends on the particle flux density.

Measuring units UAK-07 and BOI-04 provide analysis and mathematical treatment of the flow of electrical impulses coming from the detection units, the issuance of signals exceeding or not exceeding a certain threshold (using indicators "CONTROL", "CLEAN" and "DIRTY") and an indication of numerical measurement results.

Sert_RZBARadiometer RZBA-04-04M passed state acceptance tests on DSTU 3400, has a type approval certificate from 24.10.03 № UA-MI/I-1392-2003, issued by the State Committee of Ukraine for Technical Regulation and Consumer Policy and submitted to the State register of measuring instruments under the number U1784-03.

Sert_RF_RZBARadiometer RZBA-04-04M included in the State register of measuring instruments eligible for use in the Russian Federation, the certificate number 27 653 UA.C.38.999.A.

 

 

The range of flux density measurement of beta particles from the surface, min-1∙см-2;
in the energy range of beta particles, MeV
from 5 to 15000
from 0,15 to 2,5
Maximum permissible relative error in measuring the density of beta-particles flux at
P = 0.95 does not exceed,%
± 30
Measuring range of flux density of alpha particles from the surface, min-1∙см-2
in the energy range of alpha particles, MeV
from 0,1 to 10000
from 4,13 to 5,6
Maximum permissible relative error of measurement radiometer flux density of alpha particles from the surface at P = 0.95 does not exceed,% ± 30
The sensitivity of detection units BDZB to beta radiation from sources such as working standard of CO with radionuclides 90Sr+90Y no less, (imp/s)/(min-1∙см-2):
- for channels №1¸13, 16-19;
- for channels №14 и №15


0,5
0,3
The sensitivity of the detection unit BDZA to alpha radiation from sources such as impact reference P9 with radionuclide 239Pu no less, (imp/s)/(min-1∙см-2) 0,2
The sensitivity of the detection unit BDZA to alpha radiation from sources such as impact reference U8 with radionuclide 238U no less, (imp/s)/(min-1∙см-2). 0,08
The sensitivity of the detection unit BDZA to alpha radiation from sources such as impact reference U4 with radionuclide 234U no less, (imp/s)/(min-1∙см-2) 0,04
Thresholds alarms are set individually for each measuring channel, min-1∙см-2:
- for the detection unit BDZA within the limits;
- for the detection unit BDZB within the limits.

from 0,1 to 9999
from 5 to 9999
Setting resolution thresholds, min-1∙см-2:
- at threshold levels of less than 5 min-1∙см-2;
- at threshold levels of more than 5 min-1∙см-2

0,1
1
Time of continuous work, hours per day no limit
Operation mode setup time by no more, min 2
The instability of the radiometer readings for continuous operation 24 hours max,% 5
Power consumption in view of the power consumed by the analyzer connected to the units and devices detect no more, AB 150
MTBF, hours 10000
Average lifetime no less, years 10

Permissible operating conditions

Environment temperature, °С from +5 to +50
Relative humidity (at 35°С), % 95
Barometric pressure, kPa from 84 to 106,7

Dimensions and weight depend on the composition of radiometer performance.
  Composition, weight and dimensions listed in the file RZBA.pdf (Download)

Radiometer surface contamination of alpha-and beta-active substances RZBA-04-04M manufactured in various designs. Depending on the performance of the radiometer can be:
– different number of channels and, accordingly, the total area of different detection units;
– optional external probes to measure the contamination of surface alpha-and beta-active nuclides that are connected to an autonomous unit of measurement;
– stationary detection units for the control of the back of the palms;
– mobile detection unit for measuring the contamination of the upper part of head;
– embedded industrial computer;
– one or two disciplinary barrier to the passage of personnel.

Modification 20, 40 (basic version)

Радиометр РЗБА. Исполнение 20Is a rack with 18 stationary detection unit (hereinafter DU) BDZB and unit of measurement UAK-07 (located inside the installation). Basic version is equipped with a indicating unit BIC-03, which is shown as a symbolic circuit human figure is intended to indicate the pollution in each of the 18 fixed channels. Modification 20 has a display and keyboard unit BIC-02 by which we can derive the numerical results of measurements on each of the 18 stationary channels, as well as the trimming of radiometer. Basic version is equipped with a of radiometer BD for monitoring top of the head, placed on a movable axis. Position of DU for monitoring top of the head is regulated depending on the growth of human. Also present are stationary DU for monitoring back of the hand . The base model is equipped with remote DU BDZB for monitoring the "blind spots" in beta contamination. Also in the base model has remote DU BDZA for monitoring sites of the human body by alpha contamination. The measurements results for each of the remote channel is processed and displayed on the LCD display of measurement unit BOI-04. In case of excess contamination above the established threshold for block BOI-04 red LED "DIRTY" lights up  and the audio alarm. In other case of green LED "CLEAN" lights up. Modification 20 has two disciplinary barrier for the passage of personnel. Modification 40 distinguished by the presence of only one disciplinary barrier.

Modification 20.01, 40.01

Radiometer RZBA-04-04M modification 20.01 and 40.01 are analogues of 20 and 40 modifications respectively except for one difference. Display unit BIC-03 replaced by a remote industry computer. Исполнение 20.01 имеет два дисциплинарных барьера для прохода персонала.Modification 40.01 has only one disciplinary barrier.

Modification 21,41

RZBA_21thmbRack with 15 stationary detection unit BDZB and unit of measurement UAK-07 (located inside the installation). Unlike the base model, no stationary detection units for monitoring the top of the head and back of the hand. Modification 21 has two disciplinary barriers, and the modification 41 has only one.

 

 

 

 

Modification 22,42

RZBA_22thmbRack with 15 stationary detection unit BDZB and unit of measurement UAK-07 (located inside the installation). Unlike the base model, no: stationary detection units for monitoring the top of the head and back of the hand; remote probes for monitoring the "blind" spots in beta-contamination; remote unit for monitoring parts of the body for alpha contamination; block BOI-04. Modification 22 has two disciplinary barriers, and the modification 42 has only one.

 

 

 

Modification 24,44

RZBA_24Rack with 15 stationary detection unit BDZB and unit of measurement UAK-07 (located inside the installation). Unlike the base model, no: stationary detection units for monitoring the top of the head and back of the hand; remote unit for monitoring parts of the body for alpha contamination. Modification 24 has two disciplinary barriers, and the modification 44 has only one.

 

 

 

Modification 28,48

RZBA_28Rack with 18 stationary detection unit BDZB and unit of measurement UAK-07 (located inside the installation). Unlike the base model, no remote unit for monitoring parts of the body for alpha contamination. All other nodes are similar to the basic model. Modification 28 has two disciplinary barriers, and the modification 48 has only one.

 

 

 

 

Modification 29,49

RZBA_29thmbRack with 18 stationary detection unit BDZB and unit of measurement UAK-07 (located inside the installation). Unlike the base model, no remote probes for monitoring the "blind" spots in beta-contamination. All other nodes are similar to the basic model. Modification 29 has two disciplinary barriers, and the modification 49 has only one.

 

 

 

 

Modification 32,52

RZBA_32thmbRack with 18 stationary detection unit BDZB and unit of measurement UAK-07 (located inside the installation). Unlike the base model, no: remote DU for monitoring the "blind" spots in beta-contamination; remote unit for monitoring parts of the body for alpha contamination; block BOI-04. All other nodes are similar to the basic model. modification 32 has two disciplinary barriers, and the modification 52 has only one.

 

 

 

Modification 35,55

RZBA_35thmbRack with 17 stationary detection unit BDZB and unit of measurement UAK-07 (located inside the installation). Unlike the base model, no: stationary detection units for monitoring the top of the head; remote DU for monitoring parts of the body for alpha contamination. All other nodes are similar to the basic model. modification 35 has two disciplinary barriers, and the modification 55 has only one.

 

 

 

Modification 68

RZBA_68-1thmbRadiometer RZBA-04-04M modification 68 controls contamination of surfaces of 26 parts of the body by 26 stationary units, each with a separate measuring channel. The radiometer is a stationary device, consisting of: rack with 26 stationary detection unit BDZB (BDZB-15 and BDZB-16) and unit of measurement UAK-07 (which is located inside the rack of radiometer); remote detection unit BDZB (BDZB-05) for measuring the level of contamination by beta-active substances in any part of the body surface; remote detection unit for measuring BDZA pollution levels of alpha-active substances in any part of the body surface; measurement unit BOI-04, which processes and displays the results of measurements from the remote units and BDZB BDZA on LCD; two doors with electromagnetic locks, which prevents the possibility of passing of personnel without any control. Radiometer provides audible and visual alarm. The thresholds alarms are set separately for each channel.

Modification 05

05thmbIs a separate remote for monitoring unit BDZB sections of the human body for beta contamination. The measurement results are displayed on the unit of measurement BOI-04, functions of radiometer the same as a remote detection unit BDZB on the base model of radiometer. The radiometer is equipped with a device for mounting on the wall. Unlike the basic model: there are no stationary detection units and remote detection unit for alpha contamination.

Modification 06

06thmbConsists of two remote detection unit BDZB and BDZA for monitoring sites of the human body by beta-and alpha-contamination. he measurement results are displayed on the unit of measurement BOI-04, functions of radiometer the same as a remote detection unit BDZB on the base model of radiometer. The radiometer is equipped with a device for mounting on the wall. Unlike the basic model: there are no stationary detection units.

Modification 07

07thmbIt is the stationary plant for a foot monitoring. The structure is a metal turnstile and two stationary detection units for a foot monitoring. The results of measuring had representing on a LCD of the processing block BOI-04. The monitor is equipped with the indicating elements "READY" and "NOT READY". There are not 16 stationary detection units for monitoring sections of a human body how at the base model. There are not remote detection units for monitoring sections of a body on an alpha- and a beta-pollution.

Modification 10

10thmbThe modification 10 is a separate remote unit BDZA for monitoring sections of a human body on an alpha-pollution. The results of measuring has deduced on the processing block of the information BOI-04. The functions of the monitor are analogous to functions of the remote detection unit BDZA on base model of the monitor. The monitor is equipped with adaptation for fastening on a wall. There are not the stationary detection units and the remote detection unit on a beta-pollution.

Modification 11

11thmbIt is the stationary plant for monitoring impurity of arms. The monitor is equipped with two stationary detection units BDZB and the processing block of the information BOI-04 for an output of observed data and indications of overflow fixed threshold value of pollution. The radiation monitor is equipped with adaptation for fastening on a wall. There are not 16 stationary detection units BDZB, the remote detection units for monitoring sections of a human body on an alpha- and a beta-pollution.

Modification 19

19Stationary installation for pollution control of hands and sites of the human body for beta contamination. The radiometer is equipped with two stationary detection unit BDZB and remote unit BDZB and unit of measurement UAK-07 for output and display of measurement results exceeded or not exceeded the established threshold of contamination. The radiometer is equipped with a device for mounting on the wall. Differences from basic variant: number of stationary detection units BDZB reduced to two, there is no remote detection unit for monitoring sites of the human body for alpha contamination.