IKARthmb

Radiometer IKAR is a portable Radiometer with digital display readings.

Radiometer intended for measuring volume activity of radon-222 in the air and control of temperature, pressure and humidity.

The radiometer is used by radiation monitoring services in industry and medicine.

The radiometer is recommended for:

- measuring of volume activity of radon in exploited industrial buildings, residential buildings as well as commissioning new buildings and structures;

- measuring of volume activity of radon in the ground the alleged gas output during the various types of work;

- depth study of patterns of receipt of radon in indoor air in various geological conditions, including long-term systematic observations of typical buildings;

- identifying areas of radon income (for localization).

As the detector in radiometer using an ionization chamber.

Radiometer is a portable device consisting of a measuring device YI-02 and the power supply.

Measuring device YI-02 is produced in the form of a monoblock with a handle. The structure of the measuring device YI-02 includes the following components of nodes:
- detector (ionization chamber IC-01);
- pumping node UN-01;
- amplifier unit PUI-08;
- information processing unit POI-36;
- indicating node PIC-29;
- key node PKK-12;
- battery node UA-01;
- corps.

Measurement range of volume activity of radon-222 in the air, Bq/m3 from 5 to 1×105
Operation mode setup time radiometer, min not more 20
The radiometer's continuous operation time with power from the mains supply, hours per day not less 24
The radiometer's continuous operation time when powered from internal battery (without recharging), h not less 48
The instability of the radiometer measurement results for continuous operation 24 hours, % not more 10
Radiometer is powered by AC mains voltage, V
frequency, Hz
or from internal battery
voltage, V
and electric charge, AH
from 187 to 242
50 ± 1

6
7,2
The power consumed by the radiometer from the mains, VA not more 30,0
Mean time to failure, h not less 25000
Average lifetime (including the replacement of individual modules and units, worn out), years not less 10
The volume of the detector, l
general
operating

0,62
0,56

Operating conditions

Environment temperature, °С from -20 to +50
Relative humidity (at + 35 °С), % up to 95
Barometric pressure, kPa from 54 to 540