RKS-02 "Kordon"RZBA-04-04MAMP-12 "Puls 1A"MKS-2001Pump blocks
Radiometer RUG-2001BDMG-04RW systemRadon monitoringRadiation monitoring

Radiometer RUG-2001

Radiometer RUG-2001

Main

News

About us

Production

Services

Projects

 

 

1. Destination of radiometer
2. Structure of radiometer

3. Characteristics of radiometer

To top
1. Destination of radiometer
Radiometer of specific activity radioactive nuclides RUG-2001 is intends for definition specific activity of natural radioactive nuclides 226Ra, 232Th, 40K and technogenic scales - radioactive nuclides 137Cs, 60Co in steel.
 The radiometer has applied to definition of sample steel's specific activity in the laboratories of the radiological surveillance and does not demand special preparation for a sample. The gamma scale - spectrometric method is in a base of the radiometer work and the procedure used for a measuring of a pulse-height distribution from researched radioactive nuclides in fixed energy intervals.
 The radiometer has labelled to the measuring instrument.


2. Structure of radiometer
The radiometer is a stationary arrangement. It is consisting of the following crucial components:
- the radiometric unit  with lead protection against agency of an environing background on process of measuring. Depth of lead protection in  measuring zone is 10 cm;
- the detection unit on basis NaJ (Tl) with the intrinsic well;
- the amplitude transducer unit;
- PC;
- the package of exemplary volumetric measures of activity are using for a testing of the radiometer technical condition and a checking.

To top

The radiometric unit with lead protection

Detection unit with the intrinsic well

 The radiometric unit with lead protection Detection unit with the intrinsic well

Unit of the transducer of voltages

Package of exemplary volumetric standards

The transducer unit of voltages Package of exemplary volumetric standards


 

The detection unit will transform a stream of gamma-ray energies to pulses consecution. In a butt of the detection unit is located the well for seating analyzed sample.
 The four of the detection unit could be connect simultaneously to the PC.
 Maximum distance between the PC and the detection unit is 10 m.
 The amplitude transducer is transformed the pulses amplitude from the detection unit to numeric codes and transmits them to a PC. The amplitude transducer has two independent ports of information processing; it can be simultaneously connected two detection unit. The transducer has positioned in the loose PCI-slot of a PC.
 The information exchange has effected through a COM-port.

 

SoftwareSoftware (Windows 98)
- the user interface which is including the packages of steering by technology of measuring in automatic and the conversational mode, management of a database, realization of calibration, checking and submission of the information;
- is a possibility of deriving and information processing simultaneously from four radiometric clusters;
- is a possibility of a measuring mode and an automatic calibration;
- is a possibility of documentation result by the way protocols of the selected form. the user interface including packages of control by technology of measuring in self-acting and the dialog mode, supports of a database, carrying out of sizing, checking and performance of the information;
- a possibility of deriving and information processing simultaneously from four radiometric clusters;
- control of a regime of measuring and self-acting sizing;
- a possibility of documenting of outcomes of analysis as protocols of the selected form.

To top
3. Characteristics of radiometer
 

Minimum rectified specific activity for radioactive nuclides 226Ra, 232Th, 137Cs, 60Co no more, Bk/kg 20
Minimum rectified specific activity for radioactive nuclide 40K no more, Бк/kg 100
Maximal time of measuring of one assay no more, minutes 60
The bias of specific activities does not exceed, % 25
Processing method of analog signals of detection unit spectrometric
Operating mode uninterrupted
Stabilization time of operating conditions of radiometer no more than, min 30
Supply voltage from a single-phase net alternating-current, V (frequency, Hz) 187...242 (50±1)
Power consumption, Va 250
The dimension of analyzed standard tests began across diameter, mm (in height, mm) 37±1 (25±1)


 

Possible conditions of operation
Temperatures of an ambient air, °C +10...+50
A relative humidity at temperature +35°C up, % 80


 

Dimensions
The radiometric unit 480х480х670mm, weight – 200kg
The transducer unit of voltages 115х180mm


To top



 