1. Destination and characteristics of radiometer
2.
Dosimeter-radiometer at objects
To top
1. Destination
and characteristics of radiometer
Dosimeter-radiometer MKS-2001 is the multifunction fixed instrument
with numeral indication of statement, a microprocessor control and
presence of removable detection units (BDMG-04, BDMG-100, BDMG-101,
BDMN-100, and UDZhG-01). The detection units have connected to the block
of measuring BOI-12 by the screened cable. The block of signal system
BIC-17 has delivered for a customer order. It have intended for backup
light and audible signals about overflow by the measured meaning of
fixed thresholds. The dosimeter - radiometer is applied to realization
of monitoring in establishments where are works with radiomaterials and
other radiation - hazardous sources of ionizing radiations.
|
 |
 |
|
The block of measuring
BOI-12 |
The block of signal system BIC-17 |
Dosimeter - radiometer does allow to measure
power of an equivalent a gamma-ray dose zones with low (BDMG-04, BDMG-100) and
high (BDMG-101) dosage rates, and power of an equivalent a neutron radiance dose
(BDMN-100) and volumetric activity of scales - emanating of nuclides in fluid
(UDZhG-01).
|
Block of detecting BDMG-04 |
 |
MKS-2001 with use
of the detection unit BDMG-04 has intended for measuring power of the
equivalent a gamma-ray dose in a low value of dose power zones.
-
Range of logged energies of gamma
radiation from 0,05 up to 3,0 MeV.
-
Range of measuring of power of the equivalent dose of a gamma
radiation from 0,05 Sv/h up to 10 Sv/h.
-
Limits of a supposed
intrinsic error of measuring MEqD at a fiducial probability 0,95 in
relation to standard calibration performance ± (15 + 1/ah) %.
-
Responsivity of the
detection unit at graduation on exemplary radioactive nuclide 137Cs,
not less than 4,0 s-1 / (Sv/h
∙ h-1) - BDMG-04 (the sensing channel); - BDMG-04
(the rough channel) 8,0 s-1 / (Sv/h∙
h-1); - BDMG-04-01 4,0 s-1 / (Sv/h
∙ h -1); - BDMG-04-02 8,0 s-1 / (Sv/h
∙ h-1); - BDMG-04-03 0,3 s-1 / (Sv/h
∙ h-1).
|
|
Presentation of detection
unit BDMG-04 (PowerPoint) |
|
|
|
Block of detecting BDMG-100 |
 |
MKS-2001 with use
of the detection unit BDMG-100 has intended for measuring power of the
equivalent gamma-ray dose in a low value of dose power zones.
-
Range of filed energies
of a gamma radiation from 0,05 up to 3,0
МeV.
-
Range of measuring of power of the equivalent dose of a gamma radiation
from 0,1 Sv/h up to 10 Sv/h.
-
Limens of a supposed
basic relative error of measuring an equivalent power dose of a
gamma-ray at a confidence coefficient 0,95 aren't more ± 25 %.
-
Sensitivity of the detection unit
to a gamma-ray is: not less than 3,9 imp/s
on 1 Sv/h on the sensitive channel and not less than 4,9 impulse/s on 1
Sv/h on the rough channel.
|
|
Block of detecting BDMG-101 |
 |
MKS-2001 with use of the detection unit BDMG-101 has intended for
measuring power of the equivalent gamma-ray dose in high value of dose power
zones.
-
Range of filed energies
of a gamma radiation from 0,05 up to 3,0
МeV.
-
Range of measuring of
power of the equivalent dose of a gamma radiation from 3x10-4
Sv/h up to 200 Sv/h
-
Limits of an assumptive
basic relative miscalculation of measurement an equivalent power
dose gamma-ray at confidential probability 0,95 aren't more ± 25 %
-
Change of sensitivity at
permanent an equivalent dose power in dependence from
energy of registered radiation in a range from 8 up to 480
fJoule (from 0,05 up to
3,0 МeV) aren't more than 30 % from the value received from an exemplary
source cesium - 137 (106 fJoule)
|
|
Block of detecting BDMN-100 |
 |
MKS-2001 with use of the detection unit BDMN-100 has intended for
measuring an equivalent dose power of the neutron radiation.
-
Range of filed energies
of a neutron radiation
from 0,025 up to 10,0 МeV.
-
Range of measuring of
power of the equivalent dose of a neutron radiation from 0,1
Sv/h up to 100 Sv/h.
-
Limits of an assumptive basic relative miscalculation of
measurement an equivalent dose power of a neutron radiation at
confidential probability 0,95 no more ± 25 %.
-
Sensitivity of the
detection unit to radiation
of Рu-Ве source isn't less than 0,45 imp/s on 1 Sv/h.
|
|
The device of
detection UDZhG-01 |
 |
MKS-2001 with use of the device of detection UDZhG -01 has intended for
measuring volumetric activity - gamma radiating of nuclides in a fluid.
-
Range
of filed energies of gamma-quantum's from 0,05…3,0 МeV.
-
Range of measuring of
volumetric activity of nuclides of a fluid (on a gamma radiation) from
4,0x104…4,0x109
Bq∙м-3.
-
Limits of an admitted basic relative accuracy of measuring the volumetric
solution activity of a radionuclide 137Cs at confidence probability
0,95 aren't more ±35 %.
|
To top
MKS-2001 has issue information as a digital kind on the LCD display,
light and an acoustic alarm, a communication of information on two
independent standard interfacings RS-232 and RS-485.
The dosimeter - radiometer writes down in a memory: the background
values, the overflow of a threshold the information, the facts of
turning power the dosimeter - radiometer for monitoring behind the
unauthorized circulation. The information noted in memory could not
be removed of the operator.
Overall dimensions and weight
of the component parts
|
Appellation of the
component part |
Overall dimensions,
mm, no more |
Weight,
kg, no more |
|
width |
height |
depth |
|
Block of detecting BDMG-101:
-module of collision chamber (without
cable)
-module of electrometer (without cable) |
diameter – 90, length –
210
diameter – 90, length – 210 |
1,5
1,5 |
|
Block of detecting BDMG-04 |
diameter - 50, length - 400 |
0,8 |
|
Block of detecting BDMG-100 |
50 |
230 |
60 |
0,4 |
|
Block of detecting BDMN-100 (with
bracket and inhibitor) |
270 |
390 |
270 |
12,0 |
|
Block of
detecting UDZhG -01 |
1100 |
850 |
790 |
1120,0 |
|
The block of measuring
BOI-12 |
200 |
570 |
120 |
4,8 |
|
The block of signal system
BIC-17 |
160 |
400 |
100 |
2,5 |
The
dosimeter-radiometer preserves the performances under following conditions
exploitation:
-
An ambient temperature from a plus 5 0С
up to a plus 50 0С
for BOI-12 and BIC-17;
- An ambient temperature from a minus 30 0С
up to a plus 50 0С
for BDMN-100;
- A velocity of change of temperature -than more 5 0С/h;
- A relative humidity - up to 75 % (the upper value) at temperature + 50 0С
at a long-term environment and the limiting relative humidity (95 ± 3) % at
temperature + 50 0С
at short-lived action during 2 hour;
- Air pressure does from 54 up to 540 kPa.
Limits of the supposed variation of the mean error caused by change of
temperature, no more ± 5 % on everyone 10 0С
rejections from 20 0С.
2. Dosimeter-radiometer at
objects
- The detached subdivision Rovno NPP, Kuznetsovsk 4 units since 2004;
- The detached subdivision Khmelnitskiy NPP, Netishin 8 units since 2004;
- Chernobyl NPP, Chernobyl 6 units since 2004;
- HOIT-2 GSP Chernobyl NPP, Chernobyl 3 units since 2004;
- The detached subdivision Zaporozhye NPP, Energodar 2 units since 2004.
To top
|