The diagrammatic representation of detection unit BDMG-04
connection (1)
of through a connecting box KRS-29 (2) to
the measuring unit of dosimeter - radiometer MKS-2001 (3) |
Detection unit has a built-in microcontroller which managing of its
devices and had determined of gauged value. Detection unit can
work in handing over regime of the information on gauged value as
subsequent pulses and in handing over regime of the information on
gauged value at interfacing RS-485.
The microcontroller of the detection unit to evaluation of gauged
value on a base of a gauged counting rate from Geiger-Muller
counters with the count of a statistical property of allocation
pulses at time. The microcontroller has linearization of a plateau
counters characteristic of mathematical methods too. The
microcontroller monitoring a high voltage on counters and fulfills
monitoring procedures of work capacity the detection unit.
3. Preferences
1.
The detection unit is fasten to a vertical surface
(to a wall, a girder, a design, etc.) with help a holder. The
design of the holder does ensure a possibility of fast changing the
detection unit on a plant. That is especially actually at the
detection unit determination in heightened level MEqD of a gamma ray
zones. Fastening of the detection unit in a holder is kind of with
the help of a tilting yoke. The tilting yoke has positioned into
work position with the help of a nut, which has fixed without
application of the special instrument. The holder fastens on a wall
with the help of expansion bolts with the screws, included in a
package of detection unit delivery.
|
 |
|
Design of detection unit BDMG-04 holder
(1 - holder; 2- tilting yoke) |
To top
2. For conveniences of detection unit
wiring on a plant, the connecting box KRS-29 to be part of package
delivery.
|
 |
|
Connecting box KRS-29 |
Connecting box KRS-29 is ensuring
of communication cables connections to spring terminal connectors
without unbrazing of scutoff points on a plant.
There are set the elements in the connecting box which providing
realization of requirements to electromagnetic compatibility (has
ensuring normal operation of the detection unit in requirements of an
electromagnetic environment of average stiffness, group of fulfillment
on noise stability П3 on НП 306.5.02/3.035).
The connecting box has a demountable cover, which fastens to the cage of
a box with the help of four screws.
On an interior of connecting box cover is the table of connection with
notation and the name of contact pieces for connection BDMG-04 to
equipments.
 |
 |
Cover of connecting box KRS-29
with table of connection |
Spring terminal connectors
for connection of the communication cable |
Connecting box KRS-29 has fastened on a wall with the help of screws,
consisting of delivery package the detection unit.
Connection of the detection unit to the connecting box is kind of by
cable to form of KRS-29. Length of a cable – 1,5 m. At the advanced
order, the length of a cable can be enlarged up to 5 m.
To top
3. The gaiter does protective of connection from action of external
factors is positioned on a cutoff point bridging a cable of a connecting
box with the detection unit.
|
 |
|
The gaiter of
cutoff point
(1) |
4. For a possibility connection of the detection unit to anyone
low-voltage direct-current power supply on plant of exploitation,
the current supply of the detection unit is ensure in a wide range.
The current supply of the detection unit has carried out from the
direct-current power supply with a strain from +8,0V up to +42,0V.
The inner transducer of a strain makes a strain +12V for a power
supply the schema of the detection unit and a high voltage for a
power supply of GM-tubes.
5. The detection unit can give away the information as an
impulse sequence and on interfacing RS-485.
6. The broad bandwidth of measuring and
presence of the detection unit modifications are allowing the
optimally for select the necessary detection unit in dependence on
problems and requirements of its application..
4. Modification of detection unit
The detection unit has five modifications:
|
 |
 |
 |
 |
 |
|
Detection unit
BDMG-04-00 |
Detection unit
BDMG-04-02 |
Detection unit
BDMG-04-01 |
Detection unit
BDMG-04-04 |
Detection unit
BDMG-04-03 |
The
GM-tubes in one's capacity of the sensors had used in the detection
units.
Detection units BDMG-04-02, BDMG-04-03 and BDMG-04-04 have one
GM-tube. The detection units BDMG-04-00 and BDMG-04-01 have two
GM-tubes. Each of GM-tubes is part of a separate measuring channel
(rough and responsive for BDMG-04-00 and rough and average
sensitivity for BDMG-04-01).
Outline dimensions of the detection unit modification are differing
only in length
To top
5. Performance attributes
Service conditions of the detection unit:
-
Ambient temperature from a minus 40 up to a plus 70 °С;
-
Velocity of change of temperature - 10 °С/h;
-
Atmospheric pressure from 54 up to 540 kPa;
-
The
upper value of relative humidity (75±3) % at temperature 70 °С and limiting
value of relative humidity up to 100 % at temperature 50 °С.
Performance of detection unit:
-
Range of
logged energies of a gamma-ray from 0,05 up to 3,0 MeV.
-
The
gauging range MEqD of a gamma-ray is given below:
|
Designation |
Measuring
range, Sv/h |
Comment |
|
BDMG-04-00 |
from 0,05×10-6
to 10 |
Two channels: №1
(with high sensitivity GM-tubes) and №2 (with low
sensitivity GM-tubes) |
|
BDMG-04-01 |
from 1×10-6
to 10 |
Two channels: №1
(with average sensitivity GM-tubes) and №2 (with low
sensitivity GM-tubes) |
|
BDMG-04-02 |
from 0,05×10-6
to 2×10-3 |
One channel (with
high sensitivity GM-tubes) |
|
BDMG-04-03 |
from 0,1×10-3
to 10 |
One channel (with
low sensitivity GM-tubes) |
|
BDMG-04-04 |
from 1×10-6 to 50×10-3
|
One channel (with
average sensitivity GM-tubes) |
δHi
=
± (15+ К/Ах) %, where: Ax -
gauged value MEqD, Sv/h;
К = 1×10-6
- for BDMG-04-00 (at MEqD less 0,1×10-3 Sv/h) and
BDMG-04-02 (at MEqD less 0,1×10-3 Sv/h);
К = 2×10-6
- for BDMG-04-00 (at MEqD less from 0,1×10-3 Sv/h
up to 2×10-3 Sv/h) and BDMG-04-02
(at MEqD from 0,1×10-3 Sv/h);
К = 2×10-5 - for BDMG-04-01 (at
MEqD up to 20×10-3 ) and BDMG-04-04;
К = 2×10-3 - for BDMG-04-00 (at MEqD more 2×10-3 Sv/h), BDMG-04-01 (at MEqD
more 20×10-3
Sv/h) and BDMG-04-03).
-
Power dependence of detection unit
sensitivity to concerning power 662 keV gamma-rays of radioactive
nuclide 137Cs over the range energies from 0,05 up to 0,66 MeV is in
limens from a minus of 25 % up to a plus of 25 %, over the range
energies from 0,66 up to 1,25 MeV - from a minus of 25 % up to a
plus of 40 %.
-
Stabilization time of an operating mode
of the detection unit no more than 2 minutes.
-
Time of continuous operation of the
detection unit not less than 24 h in a day.
-
Instability of detection unit
indications during continuous operation 24 h no more than 5 %.
-
The detection unit are stable against
action of a gamma-ray with absorbed dose rate of a gamma-ray up to
1,5×10-4 Gray/h.
-
The design and coating materials of the
detection unit is ensuring a possibility of carrying out of a
decontamination (are stable against machining the cage by a napkin,
the wetted one of deactivating solutions of composition 5, 8, 9 or
10, the application 3 GOST 29075).
- Oxalic (or citric) an acid - 10 g / dm3;
- 5 % solution of a citric acid in alcohol С2Н5ОН (firmness of 96 %)
- for intrinsic surfaces;
- Fumaroles acid - 16 g / dm3, Na2S2O3 × 5H2O - 1 % solution;
- Synthetic scours such as "News", ОП-7, ОП-10;
-
The detection unit is falling into to
recovered equipment of the long-lived use.
-
The averages mean lifetime of the
detection unit not less than 25000 h provided that the radiation
resource of the GM-tubes will not be exceed.
-
The average life cycle of the detection
unit, not less than 10 years consideration of changing of separate
modules and the clusters which working out the resource.
-
The average time of a up state
regeneration of the detection unit at use of reserve parts from
composition spare parts no more than 2 h.
Overall dimensions and weight
of the component parts
|
Appellation of the
component part |
Overall dimensions,
mm, no more |
Weight,
kg, no more |
|
diameter |
length |
| Detection
unit BDMG-04-00 |
50 |
420 |
0,8 |
| Detection
unit BDMG-04-00 |
50 |
320 |
0,8 |
| Detection
unit BDMG-04-00 |
50 |
360 |
0,8 |
| Detection
unit BDMG-04-00 |
50 |
230 |
0,5 |
|
Detection unit BDMG-04-00 |
50 |
200 |
0,5 |
|
Connecting box KRS-29 |
width |
height |
depth |
2,2 |
|
160 |
220 |
95 |